Hanna-instruments HI 221 Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Hanna-instruments HI 221. Hanna Instruments HI 221 User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 40
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 0
1
Instruction Manual
HI 221 HI 223
pH/mV/ºC
Bench Meters
with Calibration Check
www.hannainst.com
Manufacturer since 1978
Functions
pH Calibration
HI 223 Calibration Check
Microprocessor pH Meter
Confirm
2
Buffer
nd
CFM
Select 2
Buffer
nd
/
Confirm
1 Buffer
st
CFM
/
Select 1
Buffer
st
Start Calibration
CAL
CFM
/GLP
Show GLP Information
Set Instrument Parameters
Clear Log/Cal Information
SET/CLR
RCL
Recall Log Information
LOG
Store Log Information
Select pH/ORP
RANGE
RCL
LOG
CLR
SET
C
C
RANGE
GLP
CFM
CAL
C
ATC
pH
SLOW
FAST
POOR
GOOD
Zobrazit stránku 0
1 2 3 4 5 6 ... 39 40

Shrnutí obsahu

Strany 1 - HI 221 HI 223

1Instruction ManualHI 221 HI 223pH/mV/ºCBench Meterswith Calibration Checkwww.hannainst.comManufacturer since 1978FunctionspH CalibrationHI 223 Cal

Strany 2 - WARRANTY

10ATCPROCEDUREThe user has a choice of 7 memorized buffers: pH 1.68, 4.01, 6.86, 7.01,9.18, 10.01 and 12.45.It is always recommended to perform a two-

Strany 3 - GENERAL DESCRIPTION

110ATCCALpHCATC Press the ARROW keys to select a different buffer value, if necessary. When the reading is stable and close to the selected buffer,

Strany 4 - Rear Panel

12CALBUFFER 1pHELECTRODECLEANATCATCThe stored calibration history to used issue error and warning messages duringcalibration to help ensure the highes

Strany 5 - SPECIFICATIONS

132CALCLEAR CAL if new electrodeATCBUFFER 1pHATCCALCFMBUFFER 1pHATCELECTRODECLEANELECTRODEThis message appears if the cleaning procedure performed as

Strany 6

14When using an appropriate HANNA P Type BNC electrode with pin, HI 221and HI 223 will assess electrode condition and response time during eachcalibra

Strany 7 - OPERATIONAL GUIDE

154GLP is a set of functions that allows the storage and retrieval of data regardingthe maintenance and status of the electrode.All data regarding the

Strany 8 - POOR GOOD

16_ The date (mm.dd)._ The year (yyyy)._ The pH calibration offset._ The pH calibration slope in mV/pH normalized to 25 °C.Note: If you calibrate usin

Strany 9

176_ The selected resolution of the instrument during calibration (HI 223 only)._ The calibration Alarm Time-Out status: if disabled o

Strany 10 - CLEAR CAL if new electrode

18Up to 100 (HI 221) or 500 (HI 223) LOG samples can be stored in memory.LOGGING THE CURRENT DATATo store the current reading into memory press the

Strany 11

198Otherwise the instrument will display log data on the primary LCD and thelast stored record number on the secondary LCD, along with the LOG andth

Strany 12 - ERROR MESSAGES

2TABLE OF CONTENTSWARRANTY ... 2PREL

Strany 13 - WARNING MESSAGES

20The date on the primary LCD and the dAtE message on the secondary LCD.The year on the primary LCD and the YEAr message on the secondary LC

Strany 14 - ELECTRODE RESPONSE TIME

210The dEL message on the primary LCD and the selected record or date on the secondary LCD. The CFM and the DEL tags will blink:Press SET key

Strany 15

22Setup mode allows viewing and modification of the following parameters: Calibration Alarm Time Out Instrument ID Current Time (hour & minute)

Strany 16

232Press the ARROWkeys to change the instrument ID value.Press the CFM key to save the modified instrument ID value or press the CALkey to cancel wit

Strany 17

24CURRENT YEAR SETPress the CAL key when the current year is displayed. The year will beginblinking.Press the ARROWkeys to change the year.Press the

Strany 18

254The instruments are factory calibrated for temperature.Hannas temperature probes are interchangeable and temperature calibra-tion is not necessary

Strany 19

26CALCcontaining ice and water. Wait a few minutes for the probe to stabilize. Press the CAL and LOG keys simultaneously and then switch the instru-m

Strany 20

276All instruments are factory calibrated for mV.Hannas pH and ORP electrodes are interchangeable and mV calibration isnot needed when replaced.If th

Strany 21

28Data transmission from the instrument to the PC can be done with the HI92000 Windows® compatible software (optional). The HI 92000 also offersgraph

Strany 22

298SIMPLE COMMANDSRNG Is equivalent to pressing the RANGE keyCAL Is equivalent to pressing the CAL keyCFM Is equivalent to pressing the CFM/GLP keyUP

Strany 23

3PRELIMINARY EXAMINATIONRemove the instrument from the packing material and examine it carefully tomake sure that no damage has occurred during shipp

Strany 24

30Temperature has an effect on pH. The calibration buffer solutions are affectedby temperature changes to a lesser degree than normal solutions. Durin

Strany 25 - TEMPERATURE CALIBRATION

310PREPARATIONRemove the protective cap.DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal withelectrodes. They will disappear when rinsed

Strany 26

32For Refillable Electrodes:If the filling solution (electrolyte) is more than 2½ cm (1") below the fill hole,add HI 7082 or HI 8082 3.5M KCl El

Strany 27

332IMPORTANT: After performing any of the cleaning procedures, rinse theelectrode thoroughly with distilled water, refill the reference chamber withfr

Strany 28 - PC INTERFACE

34The resistance of glass electrodes partially depends on the temperature. Thelower the temperature, the higher the resistance. It takes more time for

Strany 29

3541.0 Mol L-1 Na+0.1 Mol L-1 Na+Sodium Ion Correction for the Glass at 20-25 °CConcentration pH Error13.0013.5014.0012.5013.0013.5014.000.100.140.2

Strany 30 - TEMPERATURES

36pH CALIBRATION SOLUTIONSHI 70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcsHI 70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcsHI 70010P pH 10.01 Buffer Sach

Strany 31 - & MAINTENANCE

376120 mm4.7"12 mm0.5""S" VERSIONHI 1053HI 8072 1M KNO3 Electrolyte in FDA approved bottle, 4x30 mLHI 8082 3.5M KCl Electrolyte in

Strany 32

38150 mm5.9"12 mm0.5""S" VERSIONHI 3131HI1083P; Use: biotechnology, micro titration.Glass-body, micro, Viscolene, nonrefillable,

Strany 33 - TROUBLESHOOTING GUIDE

398Before using this product, make sure that it is entirely suitable for theenvironment in which it is used.Operation of this instrument in residenti

Strany 34 - FOR pH SENSITIVE GLASS

4FUNCTIONAL DESCRIPTION1) Liquid Crystal Display (LCD).2) CFM / GLP key, to confirm different values or to display Good LaboratoryPractice informat

Strany 35

40MAN22304/04 R0Australia:Tel. (03) 9769.0666  Fax (03) 9769.0699China:Tel. (10) 88570068  Fax (10) 88570060Egypt:Tel. & Fax (02) 2758.683German

Strany 36 - REFILL ELECTROLYTE SOLUTIONS

5HI 221SPECIFICATIONSegnaRHp00.61ot00.2Vm9.996±Vm0002±C°0.021ot0.02noituloseRHp10.0)Vm9.996±(Vm1.0)Vm0002±(Vm1C°1.0ycaruccAF°86/C°[email protected]±)Vm9.99

Strany 37 - ORP PRETREATMENT SOLUTIONS

6HI 223SPECIFICATIONSegnaR00.61ot00.2Hp000.61ot000.2Hp9.999±Vm0002±Vm°0.021ot0.02CnoituloseR10.0Hp100.0Hp1.0Vm9.999±()Vm1Vm0002±()Vm°1.0CycaruccAF

Strany 38 - OTHER ACCESSORIES

7OPERATIONAL GUIDEPOWER CONNECTIONPlug the 12 VDC adapter into the power supply socket.Note: These instruments use non volatile memory to retain the

Strany 39 - RECOMMENDATIONS FOR USERS

8 The pH is displayed on the primary LCD and the temperature on thesecondary LCD. The pH reading is out of range, the closest full-scale value will

Strany 40 - AND TECHNICAL SERVICE

9 The instrument automatically defaults to the ORPmeasurement mode if HANNA P type ORP electrodeis detected. Submerge the ORP electrode tip (4 cm/1

Příbuzné modely HI 223

Komentáře k této Příručce

Žádné komentáře