Hanna-instruments HI 110 Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Hanna-instruments HI 110. Hanna Instruments HI 110 User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 40
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 0
1
pH/mV/°C
Bench Meters
with Custom Buffers
HI 110 HI 111
HI 112 HI 113
Instruction Manual
pH
MR
MEM
SET
C
C
RANGE
GLP
CFM
CAL
pH / ORP Meter
HI 113
Select Setup Item
Select pH/ORP
RANGE
SET
Set Parameters
GLP
Show GLP Data
MEM
Store Reading
MR
Memory Recall
Functions
Confirm Calibration Point
CFM
Select pH Buffer
/
Enter/Exit Calibration
CAL
pH Calibration
www.hannainst.com
Zobrazit stránku 0
1 2 3 4 5 6 ... 39 40

Shrnutí obsahu

Strany 1 - HI 112 HI 113

1pH/mV/°CBench Meterswith Custom BuffersHI 110 HI 111HI 112 HI 113Instruction ManualpHMRMEMSETCCRANGEGLPCFMCALpH / ORP MeterHI 113Select Setup Ite

Strany 2 - TABLE OF CONTENTS

10RELATIVE mV MEASUREMENTS (HI 111 and HI 113) Press RANGE until rEL message will be displayed on the secondary LCD linefor about one second and m

Strany 3 - GENERAL DESCRIPTION

11PREPARATIONPour small quantities of the buffer solutions into clean beakers. If possible, useplastic or glass beakers to minimize any EMC interferen

Strany 4 - HI 110 AND HI 111

12THREE- (HI 110, HI 111) / FIVE-POINT (HI 112,HI 113) CALIBRATION Immerse the pH electrode and the temperature probeapproximately 4 cm (1½) into a

Strany 5 - SPECIFICATIONS

13 If necessary, press the ARROW keys to select a different buffer value. The   tag will blink on the LCD until the reading isstable. When the re

Strany 6 - HI 112 AND HI 113

14 To clear calibration parameters for all uncalibrated buffers startingwith current buffer, press and hold down CFM, then press CAL. Ifthis procedur

Strany 7

15GOOD LABORATORY PRACTICE (GLP)GLP is a set of functions that allows storage and retrieval of data regarding themaintenance and status of the electro

Strany 8 - OPERATIONAL GUIDE

16 The pH calibration buffers in calibrating order.The first pH calibration buffer:The second pH calibration buffer:The third pH calibration buffer:N

Strany 9

17Setup mode allows viewing and modifying the following parameters: Calibration Alarm Time-Out Custom buffer 1 Custom buffer 2 One-point calibrati

Strany 10

18CALIBRATION ALARM TIME-OUT SETPress CAL when the calibration time-out is displayed. Calibration time-out(OFF or 1 to 7 days) will start blinking.Pre

Strany 11

19INSTRUMENT ID PARAMETER SET(HI 112, HI 113 only)Press CAL when the instrument ID is displayed. The instrument ID (0000 to9999) will start blinking.P

Strany 12 - BUFFER pH

2WARRANTYHI 110, HI 111, HI 112 and HI 113 are guaranteed for two years againstdefects in workmanship and materials when used for their intended purpo

Strany 13

20Press RANGE. The day will start blinking.Press the ARROW keys to change the day.Press CFM to save the modified value.Press CAL to escape without sav

Strany 14

21All the instruments are factory calibrated for temperature.Hannas temperature probes are interchangeable and no temperaturecalibration is needed wh

Strany 15

22 Use the ARROW keys to set the reading on the secondary LCD to that ofthe hot water. When the reading is stable and close to the selectedcalibrati

Strany 16

23 mV CALIBRATION(for technical personnel only)All the instruments are factory calibrated for mV.Hannas ORP electrodes are interchangeable and no mV

Strany 17

24RELATIVE mV CALIBRATION (HI 111AND HI 113) Press CAL when the instrument is in RELATIVE mV measurement mode.The mV and   tags will blink. Absol

Strany 18

25Data transmission from the instrument to the PC can be done with the HI92000 Windows® compatible software (optional). HI 92000 also offersgraphing a

Strany 19

26CHR n Change the instrument range according with the parametervalue (n): n=0 pH range/0.001 resolution n=1 pH range/0.01 resolution n=3 mV range

Strany 20

27The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normal solutions.Du

Strany 21 - TEMPERATURE CALIBRATION

28PREPARATION PROCEDURERemove the protective cap of the pH electrode.DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal withelectrodes. Th

Strany 22

29For refillable electrodes:If the filling solution (electrolyte) is more than 2½ cm (1) below the fill hole,add HI 7082 or HI 8082 3.5M KCl Electrol

Strany 23 - mV CALIBRATION

3PRELIMINARY EXAMINATIONRemove the instrument from the packing material and examine it carefully tomake sure that no damage has occurred during shippi

Strany 24 - AND HI 113)

30CLEANING PROCEDURE General Soak in Hanna HI 7061 or HI 8061 GeneralCleaning Solution for approximately ½ hour. Protein Soak in Hanna HI 7073 or HI

Strany 25

31TROUBLESHOOTING GUIDESMOTPMYS MELBORP NOITULOSevissecxe/esnoperwolS.tfird.edortceleHpytriD nipitedortceleehtkaoS1608IHro1607IHsetunim03rofnoituloseh

Strany 26

32Typical Electrode LifeAmbient Temperature 1  3 years90 ºC Less than 4 months120 ºC Less than 1 monthAlkaline ErrorHigh concentrations of sodium ion

Strany 27 - DEPENDENCE

33pH BUFFER SOLUTIONSHI70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcsHI70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcsHI70010P pH 10.01 Buffer Sachets,

Strany 28 - & MAINTENANCE

34ORP PRETREATMENT SOLUTIONSHI7091L Reducing Pretreatment Solution, 460 mLHI7092L Oxidizing Pretreatment Solution, 460 mLpH ELECTRODESAll electrodes

Strany 29

35HI1330BGlass-body, semimicro, single junction, refillable, combination pH electrode.Use: laboratory, vials.HI1331BGlass-body, semimicro, single ju

Strany 30

36120 mm4.7"12 mm0.5"FC 10075 mm2.95"6 mm0.25"FC 200120 mm4.7"12 mm0.5"FC 210120 mm4.7"12 mm0.5"9.5mm DIA0.37&

Strany 31 - TROUBLESHOOTING GUIDE

37Kynar® is registered Trademark of Pennwalt Corp.Ultem® is registered Trademark of General Electric Co.HI1413BGlass-body, single junction, flat

Strany 32 - FOR pH SENSITIVE GLASS

38EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&ap

Strany 33 - ACCESSORIES

39RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for theenvironment in which they are used.Operation of th

Strany 34 - 4.7"

41) Liquid Crystal Display (LCD).2) CFM/GLP key, to confirm different values or to display Good LaboratoryPractice information.3) CAL key, to enter an

Strany 35 - "S" VERSION

40SALES AND TECHNICAL SERVICE CONTACTSAustralia:Tel. (03) 9769.0666  Fax (03) 9769.0699China:Tel. (10) 88570068  Fax (10) 88570060Egypt:Tel. & F

Strany 36 - 0.25"

5HI 110 AND HI 111SPECIFICATIONSEGNARHp00.61ot00.2)ylno111IH,PRO&ESI(Vm9.996±)ylno111IH,PRO&ESI(Vm0002±Cº0.021ot9.9NOITULOSERHp10.0)ylno111I

Strany 37 - ORP ELECTRODES

61) Liquid Crystal Display (LCD).2) CFM/GLP key, to confirm different values or to display Good LaboratoryPractice information.3) CAL key, to enter an

Strany 38

7HI 112 AND HI 113SPECIFICATIONSEGNARHp00.61ot00.2Hp000.61ot000.2)ylno311IH,PRO&ESI(Vm9.999±)ylno311IH,PRO&ESI(Vm0002±Cº0.021ot9.9NOITULOSE

Strany 39

8OPERATIONAL GUIDEPOWER CONNECTIONPlug the 12 VDC adapter into the power supply socket.Note: These instruments use non volatile memory to retain the p

Strany 40

9If measurements are taken successively in different samples, it is recommendedto rinse the electrode thoroughly with deionized water or tap water and

Příbuzné modely HI 111 | HI 112 | HI 113 |

Komentáře k této Příručce

Žádné komentáře