Hanna-instruments HI 2216 Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Hanna-instruments HI 2216. Hanna Instruments HI 2216 User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - Bench Meter

1www.hannainst.compH/mV/ISE/°CBench Meterwith Custom Buffersand Interval LoggingHI 2216Instruction Manual

Strany 2

103 cm(1¼")The custom buffers allow the user to calibrate in a buffer solution differentfrom a standard one. Up to two custom buffers can be set

Strany 3

11• The “ ” tag will blink on the LCD until the reading is stable.• When the reading is stable and close to the selectedbuffer, the “READY” tag will b

Strany 4 - FUNCTIONAL DESCRIPTION

12• When the reading is stable and close to the selectedbuffer, the “READY” tag will be displayed and the“CFM” tag will blink.• Press CFM to confirm t

Strany 5 - SPECIFICATIONS

13Clear calibration parameters and proceed with calibrationfrom the current calibration point. The instrument will keepall confirmed values during cur

Strany 6 - OPERATIONAL GUIDE

14The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normal solutions.Du

Strany 7

15It is recommended to calibrate the instrument frequently, especially ifhigh accuracy is required.The ppm range should be recalibrated:• Whenever the

Strany 8

16• If necessary, press the ARROW keys to select a different standardvalue.• The “ ” tag will blink on the LCD until the readingis stable.• When the r

Strany 9

17Default slope value (mV/decade):–59.16 (monovalent cation) - ion charge is 159.16 (monovalent anion) - ion charge is –1–29.58 (divalent cation) - io

Strany 10 - (1¼")

18GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GL

Strany 11

19Notes: • When the instrument is not calibrated or calibration is cleared(default values loaded) there is no “expired calibration”,and the display al

Strany 12

2TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSWARRANTY ...

Strany 13

20The third pH calibration buffer:The fourth pH calibration buffer:The fifth pH calibration buffer:Notes: • The “OLd” message displayed beside the pH

Strany 14

21ISE CALIBRATION DATAThe ISE calibration data is stored automatically.To view ISE calibration data, press GLP when the instrument is in ISEmeasuremen

Strany 15

22- or the number of days until the calibration alarm will be displayed.- or if expired (7 days ago).• The instrument ID.Notes: • If a one-point calib

Strany 16

23LOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONUp to 700 LOG samples can be stored into memory.200 manually logged

Strany 17

24LOT LOGGINGSelect “StAb” (stability logging) or the desired time interval.To start interval logging press LOG key while the instrument is inmeasurem

Strany 18

25No ISE records:Otherwise, the instrument will display the lot number on the primary LCDline, the number of records on the secondary, “LOG” tag and “

Strany 19

26• The time (hh.mm.ss) is displayed on the LCD, along with the“TIME” tag.• The date (yyyy.mm.dd) is displayed on the LCD along with the“DATE” tag.• T

Strany 20

27The “dEL” message is displayed on the primary LCD and theselected record on the secondary LCD, along with ”LOG” tag.Notes: • The ARROW keys can be u

Strany 21

28Setup mode allows viewing and modifying the following parameters:• Expired Calibration Alarm• Log Interval• First Custom Buffer• Second Custom Buffe

Strany 22 - • The instrument ID

29LOG INTERVALPress CAL when log interval is displayed. The log interval and “CFM” willstart blinking (“MAn” for log on demand, “StAb” log on stabilit

Strany 23

3PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONRemove the instrument from the pac

Strany 24

30ONE-POINT CALIBRATION BEHAVIORPress CAL when “1 Pnt” message is displayed on the secondary LCD.One of the two options (“Pnt” or “OFFS”) and “CFM” wi

Strany 25

31CURRENT TIMEPress CAL when the current time is displayed. The hour and “CFM” willstart blinking.Press the ARROW keys to change the hour.Press RA

Strany 26

32INSTRUMENT IDPress CAL when the “InId“ is displayed. The instrument ID (“0000“ to“9999“) and “CFM” will start blinking.Press the ARROW keys to c

Strany 27

33All the instruments are factory calibrated for temperature.Hanna’s temperature probes are interchangeable and no temperaturecalibration is needed wh

Strany 28 - SETUPSETUP

34• Use the ARROW keys to set the reading on the secondary LCD to thatof the hot water.• When the reading is stable and close to the selectedcalibrati

Strany 29

35PC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEData transmission from the instrument to the PC can be done with theHI 92000 Windows® co

Strany 30

36CHR xx Change the instrument range according with the param-eter value (xx):• xx=00 pH range/0.001 resolution• xx=01 pH range/0.01 resolution• xx=02

Strany 31

37• Meter status (2 chars of status byte): represents a 8bit hexadecimal encoding.• 0x10 - temperature probe is connected• 0x01 - new GLP data availab

Strany 32

38• the calibration time, yymmddhhmmss (12 chars).• electrode condition, with sign (3 chars). The“-01” code means not calculated.• Rel mV calibration

Strany 33 - TEMPERATURE CALIBRATION

39LODPxxx Requests the xxxth pH record logged data.LODMxxx Requests the xxxth mV/Rel mV record logged data.LODIxxx Requests the xxxth ISE record logge

Strany 34

41) Liquid Crystal Display (LCD).2) CAL key, to enter and exit calibration mode.3) CFM/GLP key, to confirm calibration, different values or to display

Strany 35 - PC INTERFACE

40PREPARATION PROCEDURERemove the protective cap of the pH electrode.DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normalwith electrodes. Th

Strany 36

41For refillable electrodes:If the filling solution (electrolyte) is more than 2½ cm (1”) below the fillhole, add HI 7082 or HI 8082 3.5M KCl Electrol

Strany 37

42CLEANING PROCEDURE• General Soak in Hanna HI 7061 or HI 8061 GeneralCleaning Solution for approximately ½ hour.• Protein Soak in Hanna HI 7073 or HI

Strany 38

43TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDESMOTPMYS MELBORP NOITULOSevissecxe/esnopser

Strany 39

44Typical Electrode LifeAmbient Temperature 1 – 3 years90 ºC Less than 4 months120 ºC Less than 1 monthAlkaline ErrorHigh concentrations of sodium ion

Strany 40

45pH BUFFER SOLUTIONSHI 70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcsHI 70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcsHI 70010P pH 10.01 Buffer Sachets,

Strany 41

46ORP PRETREATMENT SOLUTIONSHI 7091L Reducing Pretreatment Solution, 500 mLHI 7092L Oxidizing Pretreatment Solution, 500 mLpH ELECTRODESAll electrodes

Strany 42

47HI 1330BGlass-body, semimicro, single junction, refillable, combination pH electrode.Use: laboratory, vials.HI 1331BGlass-body, semimicro, single ju

Strany 43 - TROUBLESHOOTING GUIDE

48FC 100BPlastic-body (PVDF), double junction, refillable, combination pH electrode.Use: general purpose for food industry.FC 200BPlastic-body (PVDF),

Strany 44

49HI 1413BGlass-body, single junction, flat tip, Viscolene, non-refillable, combinationpH electrode. Use: surface measurement.ORP ELECTRODESHI 3131BGl

Strany 45 - ACCESSORIES

5SPECIFICATIONSSPECIFICATIONSSPECIFICATIONSSPECIFICATIONSSPECIFICATIONSEGNARHp0.61ot0.2–Hp00.61ot00.2–Hp000.61ot000.2–)PRO(Vm9.999±)PRO(Vm0002±mpp0999

Strany 46

50EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&ap

Strany 47

51RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for theenvironment in which they are used.Operation of th

Strany 48

52Printed in EUROPE(ROMANIA)MAN2216 02/11Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Strany 49 - ORP ELECTRODES

63 cm(1¼")OPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEPOWER CONNECTIONPlug the 12 VDC adapter into the p

Strany 50

73 cm(1¼")If measurements are taken successively in different samples, it is recommendedto rinse the electrode thoroughly with deionized water or

Strany 51

8RELATIVE mV MEASUREMENTS• Press RANGE until “rEL” message and “mV” tag are displayed forone second. After one second the temperature will be displaye

Strany 52 - MAN2216 02/11

9pppppH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONCalibrate the instrument frequently, especially if high accuracy isrequired.The

Komentáře k této Příručce

Žádné komentáře