Hanna-instruments HI 2214 Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Hanna-instruments HI 2214. Hanna Instruments HI 2214 User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - HI 2214 HI 2215

1www.hannainst.compH/mV/°CBench Meterwith Custom Buffersand Interval LoggingHI 2214 HI 2215Instruction Manual

Strany 2

103 cm(1¼")PROCEDURECalibration can be performed up to five-points.For accurate measurements a three-point calibration is recommended.The calibra

Strany 3

11• The calibrated value will be displayed on the primary LCD and thesecond expected buffer value on the secondary LCD.• After the first calibration p

Strany 4 - FUNCTIONAL DESCRIPTION

12FOUR, THREE OR TWO-POINT CALIBRATION• Proceed as described in “FIVE-POINT CALIBRATION” section.• Press CAL after the fourth, third or second calibra

Strany 5 - HI 2214 SPECIFICATIONS

13Notes: • If the value measured by the instrument is not close to theselected buffer, “WRONG” “ ” and “WRONG” “ ” tagswill blink alternately. Check i

Strany 6 - HI 2215 SPECIFICATIONS

14Press the ARROW keys to select another buffer to bereplaced.Press CFM to confirm the buffer that will be replaced.Press CAL to leave calibration wit

Strany 7 - OPERATIONAL GUIDE

15The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normal solutions.Du

Strany 8

16RELATIVE RELATIVE RELATIVE RELATIVE RELATIVE mmmmmV CALIBRATIONV CALIBRATIONV CALIBRATIONV CALIBRATIONV CALIBRATION• Press CAL when the instrument i

Strany 9

17GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GL

Strany 10 - (1¼")

18• The pH calibration offset.• The pH calibration slope (the GLP slope is the average of the calibrationslopes; the percentage is referred to the ide

Strany 11

19Notes: • The “OLd” message displayed beside the pH value meansthat this buffer was not used during last calibration. Pressand hold down SETUP if you

Strany 12

2WARRANTYWARRANTYWARRANTYWARRANTYWARRANTYHI 2214 and HI 2215 are guaranteed for two years against defects inworkmanship and materials when used for th

Strany 13

20RELATIVE mV CALIBRATION DATALast Relative mV calibration data is stored automatically after a successfulcalibration.To view the Relative mV calibrat

Strany 14

21LOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONUp to 600 logged samples can be stored into memory.A 100 manually lo

Strany 15

22LOT LOGGING (HI 2215 only)Select “StAb” (for stability logging) or the desired time interval.To start interval logging press LOG key while the instr

Strany 16

23Otherwise:• For HI 2214 the instrument will display the pH or relative mV valueon the primary LCD and the last stored record number on thesecondary

Strany 17

24• The time (hh.mm.ss) is displayed on the LCD, along with the“TIME” tag.• The date (yyyy.mm.dd) is displayed on the LCD along with the“DATE” tag.• T

Strany 18

25The “dEL” message is displayed on the primary LCD and theselected record on the secondary LCD, along with ”LOG” tag.Notes: • The ARROW keys can be u

Strany 19

26SETUPSETUPSETUPSETUPSETUPSetup mode allows viewing and modifying the following parameters:• Expired Calibration Alarm• Log Interval (HI 2215 only)•

Strany 20 - • The instrument ID

27SECOND CUSTOM BUFFERPress CAL when “cb2” is displayed. The custom buffer (disabled – “no”or “0” to “16” pH) and “CFM” will start blinking.Press the

Strany 21

28CURRENT TIMEPress CAL when the current time is displayed. The hour and “CFM” willstart blinking.Press the ARROW keys to change the hour.Press RA

Strany 22

29INSTRUMENT IDPress CAL when the “InId” is displayed. The instrument ID (“0000“ to“9999“) and “CFM” will start blinking.Press the ARROW keys to c

Strany 23

3PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONRemove the instrument from the pac

Strany 24

30All the instruments are factory calibrated for temperature.Hanna’s temperature probes are interchangeable and no temperaturecalibration is needed wh

Strany 25

31• Use the ARROW keys to set the reading on the secondary LCD to thatof the hot water.• When the reading is stable and close to the selectedcalibrati

Strany 26 - SETUPSETUP

32PC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEData transmission from the instrument to the PC can be done with theHI 92000 Windows® co

Strany 27

33CHR xx Change the instrument range according with the param-eter value (xx):• xx=00 pH range/0.001 resolution• xx=01 pH range/0.01 resolution• xx=02

Strany 28

34• Reading status (2 chars): R - in range, O - overrange, U - under range. First character correspondsto the appropriate range reading. Second charac

Strany 29

35• Instrument ID (4 chars)• Calibration alarm time out for pH range (2 chars)• SETUP information (2 chars): 8 bit hexadecimalencoding.• 0x01 - beep O

Strany 30 - TEMPERATURE CALIBRATION

36PREPARATION PROCEDURERemove the protective cap of the pH electrode.DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normalwith electrodes. Th

Strany 31

37For refillable electrodes:If the filling solution (electrolyte) is more than 2½ cm (1”) below the fillhole, add HI 7082 or HI 8082 3.5M KCl Electrol

Strany 32 - PC INTERFACE

38CLEANING PROCEDURE• General Soak in Hanna HI 7061 or HI 8061 GeneralCleaning Solution for approximately ½ hour.• Protein Soak in Hanna HI 7073 or HI

Strany 33

39TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDESMOTPMYS MELBORP NOITULOSevissecxe/esnopser

Strany 34

41) Liquid Crystal Display (LCD).2) CAL key, to enter and exit calibration mode.3) CFM/GLP key, to confirm calibration, different values or to display

Strany 35

40Typical Electrode LifeAmbient Temperature 1 – 3 years90 ºC Less than 4 months120 ºC Less than 1 monthAlkaline ErrorHigh concentrations of sodium ion

Strany 36

41pH BUFFER SOLUTIONSHI 70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcsHI 70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcsHI 70010P pH 10.01 Buffer Sachets,

Strany 37

42ORP PRETREATMENT SOLUTIONSHI 7091L Reducing Pretreatment Solution, 500 mLHI 7092L Oxidizing Pretreatment Solution, 500 mLpH ELECTRODESAll electrodes

Strany 38

43HI 1330BGlass-body, semimicro, single junction, refillable, combination pH electrode.Use: laboratory, vials.HI 1331BGlass-body, semimicro, single ju

Strany 39 - TROUBLESHOOTING GUIDE

44FC 100BPlastic-body (PVDF), double junction, refillable, combination pH electrode.Use: general purpose for food industry.FC 200BPlastic-body (PVDF),

Strany 40

45HI 1413BGlass-body, single junction, flat tip, Viscolene, non-refillable, combinationpH electrode. Use: surface measurement.ORP ELECTRODESHI 3131BGl

Strany 41 - ACCESSORIES

46EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&ap

Strany 42

47RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for theenvironment in which they are used.Operation of th

Strany 43

48Printed in EUROPE(ROMANIA) MAN2215 02/11Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Strany 44

5HI 2214 SPECIFICATIONSHI 2214 SPECIFICATIONSHI 2214 SPECIFICATIONSHI 2214 SPECIFICATIONSHI 2214 SPECIFICATIONSEGNARHp0.61ot0.2–Hp00.61ot00.2–Vm9.996±

Strany 45 - ORP ELECTRODES

6HI 2215 SPECIFICATIONSHI 2215 SPECIFICATIONSHI 2215 SPECIFICATIONSHI 2215 SPECIFICATIONSHI 2215 SPECIFICATIONSEGNARHp0.61ot0.2–Hp00.61ot00.2–Hp000.61

Strany 46

73 cm(1¼")OPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEPOWER CONNECTIONPlug the 12 VDC adapter into the p

Strany 47

83 cm(1¼")If measurements are taken successively in different samples, it is recommendedto rinse the electrode thoroughly with deionized water or

Strany 48 - (ROMANIA) MAN2215 02/11

9RELATIVE mV MEASUREMENTS• Press RANGE until “rEL” message and “mV” tag are displayed forone second. After one second the temperature will be displaye

Příbuzné modely HI 2215

Komentáře k této Příručce

Žádné komentáře