Hanna-instruments HI 83214 Uživatelský manuál Strana 57

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 72
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 56
57
Phosphorus, Total
Press the Zero key. The instrument will perform a zero sequence and after a few seconds the display
will show “-0.0-”. Now the meter is zeroed and ready for measurement.
Remove the vial.
Remove the cap and add the content of one packet of HI 93758-0
Phosphorus Reagent.
Replace the cap tightly and shake gently to mix for about 2 minutes
until all the powder is completely dissolved. This is the sample.
Place the vial into the holder and push it completely down.
Press Timer and the display will show the countdown prior to the measurement and the “Reaction
Time” message. Alternatively, wait for 3 minutes and press Read.
The instrument will perform the reading. The instrument displays concentration in mg/L of phosphorus
(P).
The method detects free (orthophosphate) and condensed inorganic forms (meta-, pyro- and other
polyphosphates) of phosphates present in the sample.
Press s or t to convert the result in mg/L of phosphate (PO
4
3
-
) or mg/L of P
2
O
5
.
Zobrazit stránku 56