SPECIFICATIONS
Range 0.00 to 3.00 mg/L (as NH
3
-N)
Resolution 0.01 mg/L
Accuracy ±0.10 mg/L or ±5 % of reading @ 25 °C, whichever is greater
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92
,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code Description Q.ty/test Q.ty/set
HI 93764A-0* Reagent Vial 1 vial 25 vials
HI 93764-0 Nessler Reagent 4 drops 1 bottle
*
Reagent Vial identification
: A LR, white label.
Note: Store the unused vials in a cool and dark place.
REAGENT SET
HI 93764A-25 Reagents for 25 tests
For other accessories see page 68.
MEASUREMENT PROCEDURE
Select the
Ammonia LR
method following one of the procedures described in the “Method Selection”
section (see page 15).
Ammonia LR
AMMONIA LOW RANGE
Remove the cap from a Reagent Vial.
Add exactly 5.0 mL of sample to the vial, while keeping the
vial at a 45-degree angle.
Replace the cap and mix by inverting the vial a couple of
times. This is the blank.
Place the vial into the holder and push it completely down.