Calibration alarm time out for pH range (2 chars)
SETUP information (2 chars): 8 bit hexadecimal
encoding.
0x01 - beep ON (else OFF)
0x04 - degrees Celsius (else degrees Fahrenheit)
0x08 - Offset calibration (else Point calibration)
The number of custom buffers (1 char)
The custom buffer values, with sign and decimal
point, for each defined custom buffer (7 chars)
Log type 2 chars
01 - manual log
02 - stability lot log
03 to 14 - the coresponding interval for lot log (5 s
to 180 min)
cell constant, with sign and decimal point (6 chars)
TDS factor, with sign and decimal point (5 chars)
TC coef, with sign and decimal point (5 chars)
Reference Temperature, with sign and decimal point
(5 chars)
Temperature Compensation mode (1 char)
NSLx Requests the number of logged samples.
x = P - request for pH range
M - request for mV range
E - request for EC range
N - request for NaCl range
T - request for TDS range
LODxnnn: request the”nnn” record of the manual log on the “x” range
LODxALLff: groups all Log On Demand Records in frames of 8 records
each for the selected range
Command Parameters:
x - range (see Note)
ALL - download all records for the selected range
ff - requested frame number - first frame is labeled 01
LLsxff: requests information about all lots on the specified range, it sends
the information in frames of 10 lots each (a frame contains information
about 10 lots)
Command Parameters:
x - range (see Note)
ff - requested frame number - first frame is labeled 01