Hanna-instruments HI 2550 Uživatelský manuál Strana 46

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 64
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 45
46
CHRxx Change the instrument range according with the parameter
value (xx):
xx=00 pH range/0.001 resolution
xx=01 pH range/0.01 resolution
xx=02 pH range/0.1 resolution
xx=03 mV range
xx=04 Relative mV range
xx=06 EC range
xx=07 TDS range
xx=08 NaCl range
The instrument sends the “ACK” (6) character every time a command is
recognized and a “NAK” (21) character for invalid commands.
COMMANDS REQUIRING AN ANSWERCOMMANDS REQUIRING AN ANSWER
COMMANDS REQUIRING AN ANSWERCOMMANDS REQUIRING AN ANSWER
COMMANDS REQUIRING AN ANSWER
RAS Causes the instrument to send a complete set of readings in
according with the current range:
pH, mV and temperature reading on pH range.
mV and temperature reading on mV range.
Rel mV, absolute mV and temperature reading on
Rel mV range.
Conductivity and temperature reading on EC range.
TDS and temperature reading on TDS range.
NaCl and temperature reading on NaCl range.
The answer string contains:
Meter mode (2 chars):
00 - pH range (0.001 resolution)
01 - pH range (0.01 resolution)
02 - pH range (0.1 resolution)
• 03 - mV range
04 - Rel mV range
06 - EC range
• 07 - TDS range
• 08 - NaCl range
Meter status (2 chars of status byte): represents a 8
bit hexadecimal encoding.
0x10 - temperature probe is connected
0x01 - new GLP data available
0x02 - new SETUP parameter
Zobrazit stránku 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 63 64

Komentáře k této Příručce

Žádné komentáře