Submerse the temperature probe (or EC probe) in the second vessel
as near as possible to the calibrated thermometer. Allow a few
seconds for the probe to stabilize.
Use the ARROW keys to set the reading on the secondary LCD to that
of the hot water.
When the reading is stable and close to the selected
calibration point, “READY” tag will appear and
CFM” tag will blink.
Press CFM to confirm. The instrument memorize
calibration and restart in measurement mode.
Note: If the reading is not close to the selected calibration point,
WRONG” tag will blink. Change the temperature probe (or EC
probe) and restart calibration.