1www.hannainst.comInstruction ManualHI 96701CFree Chlorine ISM

Strany 2 - TABLE OF CONTENTS

10Under range: A blinking “0.00” indicates that the sampleabsorbs less light than the zero reference. Check the procedureand make sure you use the sam

Strany 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c

Strany 4 - ABBREVIATIONS

12MEASUREMENT PROCEDURE• Whenever the cuvette is placed into the measurement cell, itmust be dry outside, and completely free of fingerprints, oilor d

Strany 5 - PRECISION AND ACCURACY

13• When the beeper sounds briefly and theLCD displays dashes, the meter is ready.The blinking “ZERO” indicates that theinstrument needs to be zeroed

Strany 6 - PRINCIPLE OF OPERATION

14• Remove the cuvette.• Add the content of one packet ofHI 93701-0 Free Chlorine reagent.Replace the cap and shake gently for20 seconds (or 2 minutes

Strany 7 - FUNCTIONAL DESCRIPTION

15VALIDATION PROCEDUREOxidized Manganese and Chromium (positive error)Chlorine dioxide (positive error)Alkalinity above 250 mg/L CaCO3 or acidity abov

Strany 8

16• After a few seconds, the display willshow “-0.0-”. The meter is now zeroedand ready for validation.• Remove the cuvette.• Place the CAL CHECKTM St

Strany 9 - ERRORS AND WARNINGS

17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra

Strany 10

18• After a few seconds the display will show“-0.0-”. The meter is now zeroed andready for calibration. The blinking“READ” asks for reading calibratio

Strany 11

19• Press GLP/ to edit the desired year(2009-2099). If the key is kept pressed,the year number is automaticallyincreased.• When the correct year has

Strany 12 - MEASUREMENT PROCEDURE

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DE

Strany 13

20In the GLP mode, the last user calibration date can be verified and the factory calibration can berestored.To display the calibration date:• Press G

Strany 14

21• Press READ/ /TIMER to enter in thefactory calibration restore screen. Theinstrument asks for confirmation of usercalibration delete.• Press ZERO/C

Strany 15 - VALIDATION PROCEDURE

22To replace the instrument’s battery, follow the steps:• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the

Strany 16

23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen

Strany 17 - CALIBRATION PROCEDURE

24Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Strany 18

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl

Strany 19

4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96701 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m

Strany 20 - GLP mode

5SPECIFICATIONSPRECISION AND ACCURACYRange 0.00 to 5.00 mg/LResolution 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/LAccuracy ±0.03 mg/L

Strany 21 - BATTERY MANAGEMENT

6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross

Strany 22 - ACCESORIES

71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette ho

Strany 23 - WARRANTY

810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to me

Strany 24 - MAN96701 12/10

9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are outs