Hanna-instruments HI 83414 Uživatelský manuál Strana 30

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 44
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 29
30
Place the cuvette into the instrument. Align the
mark on the cuvette with the sign on the
instrument top and close the lid.
Press the Zero key. The dashes and the lamp
icon will blink on the LCD, depending on the
measurement phase. If the zeroing procedure
was successful, the display will show “-0.0-”.
SINGLE SAMPLE READ
Remove the cuvette from the instrument.
Remove the cap.
Add the content of one packet of the specific
test reagent, for:
Free Chlorine Total Chlorine
1 packet of or 1 packet of
HI 93701-0
HI 93711-0
Replace the cap and shake gently for 20 seconds
(or 2 minutes in case of seawater analysis).
Replace the cuvette into the holder and ensure
that the mark on the glass matches the mark on
the instrument top.
Close the lid.
Zobrazit stránku 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 43 44

Komentáře k této Příručce

Žádné komentáře