Hanna-instruments HI 83225 Uživatelský manuál Strana 54

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 60
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 53
54
Add the content of one packet of HI 93751-0
Indicator reagent.
Replace the cap and invert gently to mix for
1 minute (about 30 inversions).
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 5 minutes and press Read. When the timer ends the meter will perform the reading.
The instrument displays concentration in mg/L of Sulfate (SO
4
2
¯).
INTERFERENCES
Interferences may be caused by:
Calcium (as CaCO
3
) above 20000 mg/L
Chloride (as Cl
-
) above 40000 mg/L
Magnesium (as MgCO
3
) above 10000 mg/L
Silica (as SiO
2
) above 500 mg/L
Color or suspended matter in large amounts will interfere: suspended matter should be removed by previous
filtration.
Organic matter in large amounts may impede the precipitation of barium sulfate.
Sulfate