Phosphorus LR
Add the content of one packet of HI 93706B-0 Phosphorus
Reagent B (Amino Acid) to the cuvette. Replace the cap
and shake gently until completely dissolved.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 5 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L of phosphorus (P).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of phosphate (PO
4
3
¯) and phosphorus pentoxide
(P
2
O
5
).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO
3
Magnesium above 40000 mg/L as CaCO
3
Ferrous iron above 100 mg/L