EC calibration data (if available), which contains:
the number of calibrated standards (1 char)
the offset factor, with sign and decimal point
(7 chars)
the cell constant, with sign and decimal point
(7 chars)
the calibration time, yymmddhhmmss (12 chars)
standards information (for each standard)
standard value, with sign and decimal point
(7 chars).
standard unit (2 chars; 00-µS; 01-mS)
Reference Temperature with sign and decimal
point (5 chars)
Temperature Compensation mode (2 chars)
00 - no temperature compensation
01 - automatic temperature compensation
00 - manual temperature compensation
TC coeficient with sign and decimal point
(5 chars)
Na Cl Calibration data
the number of calibrated data (1 char)
salinity coeficient, with sign and decimal point
(7 chars)
Cell constant, with sign and decimal point
(7 chars)
calibration time, yymmddhhmmss (12 chars).
standard information:
Reference Temperature with and decimal point
(5 chars)
Temperature Compensation mode (2 chars)
00 - no temperature compensation
01 - automatic temperature compensation
00 - manual temperature compensation
TC coeficient with sign and decimal point
(5 chars)