value (xx):
xx=06 EC range
xx=07 TDS range
xx=08 NaCl range
The instrument sends the “ACK” (6) character every time a command is
recognized and a “NAK” (21) character for invalid commands.
COMMANDS REQUIRING AN ANSWER
RAS Causes the instrument to send a complete set of readings in
according with the current range:
Conductivity and temperature reading on EC range.
TDS and temperature reading on TDS range.
NaCl and temperature reading on NaCl range.
The answer string contains:
Meter mode (2 chars):
06 - EC range
07 - TDS range
08 - NaCl range
Meter status (2 chars of status byte): represents a 8
bit hexadecimal encoding.
0x40 - TDS unit (0=ppm, 1=g/L)
0x20 - EC unit (0=µS, 1=mS)
0x10 - temperature probe is connected
0x01 - new GLP data available
0x02 - new SETUP parameter
Reading status (1 char): R - in range, O - over
range, U - under range.
The reading (corresponding to the selected range) -
7 ASCII chars, including sign and decimal point.
Temperature reading - 7 ASCII chars, with sign and
two decimal points, always in ºC.
MDR Requests the instrument model name and firmware code.
GLP Requests the calibration data record.
The answer string contains:
GLP status (1 char): represents a 4 bit hexadecimal
encoding.
0x04 - EC calibration available
0x08 - NaCl calibration available
Zobrazit stránku 27