Hanna Instruments Thermometer HI 4221 Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měření, testování a kontrola Hanna Instruments Thermometer HI 4221. Hanna Instruments Thermometer HI 4221 User's Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - HI 4221 & HI 4222

1HI 4221 & HI 4222pH/mV/ISE/TemperatureBench MetersInstruction Manualwww.hannainst.com

Strany 2 - WARRANTY

10CHANNEL SELECTION (HI 4222 only)• Press while in Measure mode to access channelselection menu. Four available options will be displayed:Channel 1,

Strany 3 - TABLE OF CONTENTS

11DISPLAYING MODESDISPLAYING MODESDISPLAYING MODESDISPLAYING MODESDISPLAYING MODESFor each measurement mode (pH, mV, Rel mV or Ion) the following disp

Strany 4

12GraphAccessing this option, the online graph with currently logged values (pH, mV, Rel mV, or ISE vs. Seconds) willbe displayed.If there is no activ

Strany 5 - FUNCTIONAL DESCRIPTION

13When the online graph is displayed:•Use and to move the graph along X (Time) axis.• Press to access the zoom menu for Y axis. Use or for zo

Strany 6

14SYSTEM SETUPSYSTEM SETUPSYSTEM SETUPSYSTEM SETUPSYSTEM SETUPThe System Setup menu allows the user to customize the user interface, consult the meter

Strany 7

15• Use or to select the Beeper option.• Press and use or to highlight thedesired beeper status option you want to modify.• Press and use o

Strany 8 - SPECIFICATIONS

16To set the GLP Data:• Press while in Measure mode.• Press .• Use or to select the GLP Data option.• Press and use or to highlight thedesire

Strany 9 - OPERATIONAL GUIDE

17• Use or to select the Date & Time option.• Press and use or to highlight thedesired option you want to modify.• Press to confirm your

Strany 10

18LanguageThis option allows the user to choose the desired language in which all information will be displayed.To select the Language:• Press while

Strany 11 - DISPLAYING MODES

19Note: All the instruments are factory calibrated for mV and temperature. After one year following factorycalibration, the “Factory Calibration Due”

Strany 12

2HI 4221 and HI 4222 are warranted for two years against defects in workmanship and materials when usedfor their intended purpose and maintained accor

Strany 13

20The pH Setup menu allows the user to set the parameters associated with pH measurement and calibration.These parameters can be set specifically for

Strany 14 - SYSTEM SETUP

21To set one of the Temperature options:• Press while in pH Measure mode.• Press .• Use or to select the Temperature option.• Press and use or

Strany 15

221st Cal. PointTwo options are available for the 1st Cal. Point parameter: Point and Offset.If Point option is selected, the slope values adjacent to

Strany 16

23• Press to exit custom buffer edit menu. If the Saving Confirmation is enabled, press to acceptthe modified option, to escape without saving or

Strany 17

24Set Reminder PeriodIf choosing Daily or Periodic options for the Calibration Reminder, the Set Remind Period must be accessed in order to setthe ti

Strany 18

25Sample IDThis option allows the user to give to the measured samples an identification number/name. Two Sample IDoptions are available: ID Increment

Strany 19

26the user to select the desired character. It is also possible to delete the last character by positioning the cursor onthe Backspace character and p

Strany 20 - Manual Temperature

27To set the Reading Mode:• Press while in pH Measure mode.• Press .• Use or to select the Reading Mode option.• Press and use or to highlig

Strany 21 - Buffer Entry Type

28To set the Logging Data Configuration:• Press while in pH Measure mode.• Press .• Use or to select the Log option.• Press and use or to hi

Strany 22 - Edit Custom Buffers

29• Use or to select the Log option.• Press and use or to highlight theNew Lot option.• Press to generate a new manual lot. A pop-up menuwill

Strany 23

3TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSWARRANTY ...

Strany 24 - Clear Calibration

30To set the Alarm Limits:• Press while in pH Measure mode.• Press .• Use or to select the Alarm option.• Press and use or to highlight theA

Strany 25 - Edit Sample ID

31pH ResolutionAccessing this option, the desired pH resolution can be set, with one (x.x), two (x.xx) or three (x.xxx) decimals.To set the pH Resolut

Strany 26

32mVmVmVmVmV SETUP SETUP SETUP SETUP SETUPThe mV Setup menu allows the user to set the parameters associated with mV and Relative mVmeasurements. Thes

Strany 27 - Logging Data Configuration

33ISEISEISEISEISE SETUP (HI 4222 SETUP (HI 4222 SETUP (HI 4222 SETUP (HI 4222 SETUP (HI 4222 onlyonlyonlyonlyonly)))))The ISE Setup menu allows th

Strany 28

34Analyte AdditionThis method is similar to the Known Addition method, with the difference that an aliquot of sample is added to astandard of known co

Strany 29 - Alarm Limits

35• Press to confirm your selection or press to cancel operation.Note: If an ISE calibration was performed and the Temperature Compensation option i

Strany 30

36To set the Electrode Type:• Press while in ISE Measure mode.• Press .• Use or to select the Electrode Type option.• Press and use or to sel

Strany 31

37• To select the appropriate Electric Charge/Slope use or and then press . If the Ionelectric charge is None, its slope can be manually set bypres

Strany 32

38Isopotential PointThis option allows the user to edit the isopotential point of the electrode used for ion measurements. The ionselective electrodes

Strany 33

39pH CALIBRATION & MEASUREMENTSpH CALIBRATION & MEASUREMENTSpH CALIBRATION & MEASUREMENTSpH CALIBRATION & MEASUREMENTSpH CALIBRATION &

Strany 34

4GENERAL DESCRIPTIONGENERAL DESCRIPTIONGENERAL DESCRIPTIONGENERAL DESCRIPTIONGENERAL DESCRIPTIONRemove the instrument from the packing material and ex

Strany 35

40Three buffer entry types are available: Automatic, Semiautomatic and Manual Selection. The default option isManual Selection.To calibrate the instru

Strany 36 - ISE Measure

41• When in MTC mode, if pressing after enteringcalibration and while a custom pH calibration buffer isselected, a pop-up menu will be displayed on

Strany 37

42• Unrecognized buffer. Please check the buffer or the buffer list (for Semiautomatic and Automatic buffer entrytype): this message appears if the cu

Strany 38

43• To return to normal Measure mode press .Note: If the reading is out of range, “-----” will be displayed on the LCD.Outside Cal Range feature warn

Strany 39

44mV & Relative mV MEASUREmV & Relative mV MEASUREmV & Relative mV MEASUREmV & Relative mV MEASUREmV & Relative mV MEASUREMENTMENT

Strany 40

45Relative mVRelative mVRelative mVRelative mVRelative mV MEASUREMENTS MEASUREMENTS MEASUREMENTS MEASUREMENTS MEASUREMENTSTo measure the Relative mV o

Strany 41

46ISE CALIBRATION & MEASUREISE CALIBRATION & MEASUREISE CALIBRATION & MEASUREISE CALIBRATION & MEASUREISE CALIBRATION & MEASUREMEN

Strany 42

47Two standard entry types are available: ManualSelection and Custom Standard. The default option is Manual Selection.A. To calibrate the instrument u

Strany 43 - Rel mV Measure

48• If the standard solution concentration is validated, will appear on the LCD. Press to updatecalibration. The calibration point value will be ad

Strany 44

49• Wrong new slope. Please check the standard solution: this message appears if the current slopeexceeds slope window (50% to 120% of default slope f

Strany 45 - Relative mV Measure

5HI 4221 DESCRIPTIONFRONT PANELREAR PANEL1) Liquid Crystal Display (LCD)2) Main Keyboard3) USB connector4) ON/OFF switch5) Power adapter socket6) RS23

Strany 46

50• Add ISAB to the sample solution.• Submerge the Ion Selective Electrode tip and the temperatureprobe approximately 4 cm (1½”) into the sample to be

Strany 47

51• When the reading is stable, press to store the first mV reading. The second step of the methodwill be displayed on the LCD in which the user is n

Strany 48

52• Press to change the desired method parameters as indicated in the first step of the method.• Press to take the second mV reading.• When the read

Strany 49

53• If pressing , the method parameters can be changed and the sample concentration will berecalculated.• Press to start another measurement.Note: Pr

Strany 50

54ISE THEORYISE THEORYISE THEORYISE THEORYISE THEORYAn Ion Selective Electrode (ISE) is an electrochemical sensor that changes voltage with the activi

Strany 51

55The Nernst equation can be rewritten:ION SELECTIVE ANALYSIS METHODSDirect AnalysisDirect AnalysisDirect AnalysisDirect AnalysisDirect AnalysisThis m

Strany 52

56Example 1You have sulfide samples and you are adding Ag+. The reaction is:One mole sulfide sample reacts with 2 moles silver standard (f = ½).Exampl

Strany 53

57LOGGINGLOGGINGLOGGINGLOGGINGLOGGINGThis feature allows the user to log pH, mV (or ISE - HI 4222 only) measurements, together with temperatureautomat

Strany 54

58• If accessing Log History option while logging, last logged data can be visualized on the LCD (see DisplayMode section for details, page 11).• To s

Strany 55

59The “Logging”, sampling period and “AutoHold” indicators willbe displayed on the LCD.Note: While automatic logging is running, the measured unitsetu

Strany 56

6HI 4222 DESCRIPTIONFRONT PANELREAR PANEL1) Liquid Crystal Display (LCD)2) Main Keyboard3) USB connector4) ON/OFF switch5) Power adapter socket6) RS23

Strany 57 - LOGGINGLOGGING

60• To store another frozen value, press to return to normal logging mode and then again.• The records will be stored in one lot. In order to chang

Strany 58

61• Press , or to select the desired LogReport type. All logged lots for the selected Log Report typewill be displayed on the LCD.• To filter the

Strany 59

62PC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEData transmission from the instrument to the PC can be done with the HI 92000 Windows® c

Strany 60

63pppppH BUFFER TEMPERATUREH BUFFER TEMPERATUREH BUFFER TEMPERATUREH BUFFER TEMPERATUREH BUFFER TEMPERATUREDEPENDENCEDEPENDENCEDEPENDENCEDEPENDENCEDEP

Strany 61 - – HI 4222 only)

64PREPARATION PROCEDURERemove the protective cap of the pH electrode.DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with electrodes. T

Strany 62 - PC INTERFACE

65For AmpHelFor AmpHelFor AmpHelFor AmpHelFor AmpHel®®®®® electrodes: electrodes: electrodes: electrodes: electrodes:If the electrode does not respond

Strany 63

66TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDESMOTPMYSMELBORPNOITULOSevissecxe/esnopserwo

Strany 64

67TEMPERATURE CORRELATIONTEMPERATURE CORRELATIONTEMPERATURE CORRELATIONTEMPERATURE CORRELATIONTEMPERATURE CORRELATIONFOR FOR FOR FOR FOR pppppH SENSIT

Strany 65 - Oil/grease

68ACCESSORIESACCESSORIESACCESSORIESACCESSORIESACCESSORIESpH BUFFER SOLUTIONSHI 6001 pH 1.679 Buffer Solution, 500 mL bottleHI 6003 pH 3.000 Buffer Sol

Strany 66 - TROUBLESHOOTING GUIDE

69ORP PRETREATMENT SOLUTIONSHI 7020L Test Solution 200-275 mV, 500 mL bottleHI 7021L Test Solution 240 mV, 500 mL bottleHI 7022L Test Solution 470 mV,

Strany 67

7KEYBOARD DESCRIPTIONFUNCTION KEYSVIRTUAL KEYSThe upper row keys are assigned to the virtual keys placed on the bottom of the LCD, which allow you top

Strany 68 - ACCESSORIES

70HI 1131BGlass-body, single junction, refillable, combination pH electrode.Use: general purpose.HI 1330BGlass-body, semimicro, single junction, refi

Strany 69

71HI 2031BGlass-body, semimicro, conic, refillable, combination pH electrode.Use: semisolid products.HI 1332BPlastic-body (PEI®), double junction, re

Strany 70

72FC 200BPlastic-body (PVDF®), open junction, conic, Viscolene, non-refillable, combination pH electrode.Use: meat & cheese.FC 210BGlass-body, do

Strany 71

73ORP ELECTRODESHI 3131BGlass-body, refillable, combination platinum ORP electrode.Use: titration.HI 3230BPlastic-body (PEI®), gel-filled, combinatio

Strany 72

74EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&apos

Strany 73 - ORP ELECTRODES

75RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for the environment in which they are used.Operation of t

Strany 74

76SALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTSSALE

Strany 75

8SPECIFICATIONSSPECIFICATIONSSPECIFICATIONSSPECIFICATIONSSPECIFICATIONS1224IH2224IHEGNAR– Hp000.02ot000.2± Vm0.0002– 01*16–01*99.9ot01.cnoc– /Cº0.021o

Strany 76

9POWER CONNECTIONPlug the 12 Vdc adapter into the power supply socket.Notes: • These instruments use non volatile memory to retain the pH, Ion calibra

Komentáře k této Příručce

Žádné komentáře