Even though LP2000-11 covers a wide turbidity range, highly
accurate readings exceeding 40 FTU* require dilution as recom-
mended by the Standard Methods.
Use the formula below to calculate the amount of HI 93703-0 or
turbidity-free water needed for accurate dilution.
The total volume = 100 mL.
Vos = 3000 / T
where:
Vos = volume of original sample (mL) to be added to HI 93703-0
to obtain the total diluted volume of 100 mL.
T = LP2000-11 reading (exceeding 40 FTU)
E.g.: LP2000-11 reading = 200 FTU*
3000 / 200 = 15 mL (Vos)
15 mL (Vos)+85 mL (HI 93703-0) = 100 mL
At this point, take a sample of this diluted solution and measure its
turbidity.
The correct turbidity value of the original sample is derived by:
Ta= Tn x 100 mL / Vos
where:
T
a
= actual turbidity value of the original sample
T
n
= LP2000-11 reading of the diluted solution
E.g. If T
n
= 27
T
a
= 27 x 100 mL / 15 mL = 180 FTU*
Note: Any sample above 1000 FTU* will show out of range by blinking
“1000”.
* 1 FTU = 1 NTU
Note: Wipe the cuvet thoroughly with
HI93703-70 or a lint-free tissue before
inserting it into the measurement cell.
The cuvet must be completely free of
fingerprints and other oil or dirt, par-
ticularly in the area where the light goes
through (approx. the bottom 2 cm/1").
Place the cuvet into the cell and
check that the notch on the cap is
positioned securely into the groove.
The mark on the cuvet cap should point
towards the keyboard.
Press the READ key and the LCD will
display a blinking “SIP” (Sam-
pling in Process). The turbidity
value will appear after approxi-
mately 20 seconds.