1Calibration CheckpH/mV/TemperatureMetersHI 98140 HI 98150HI 98160Instruction Manualwww.hannainst.com

Strany 2

10Submerse the ORP electrode tip (3cm/1¼”)into the sample to be tested and wait a fewseconds for the reading to stabilize.Measurements within the ±699

Strany 3

11pppppH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONHanna recommends frequent calibrations, especially if high accuracy isrequired

Strany 4

12• If necessary, press the ARROW keys to select a different buffer value.• The “ ” tag will blink on the LCD until the reading is stable.• When the r

Strany 5 - HI 98140 & HI 98150

13• The calibrated value is then displayed on the primary LCD and thesecondary LCD will display the fifth expected buffer value.• After the fourth cal

Strany 6

14• If “WRONG” tag and “OLd” message are displayed onthe secondary LCD, an inconsistency exists between newand previous (old) calibration data. Clear

Strany 7 - SPECIFICATIONS

15CLEAR CALIBRATIONTo clear calibration press 2nd then CLR or CLR aloneafter entering CAL mode. The “CLr ALL” message willbe displayed on the LCD alon

Strany 8 - OPERATIONAL GUIDE

16The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normalsolutions. Du

Strany 9

17To enter Rel mV operation press 2nd and Mode from mV measurement.• Place the ORP electrode into solution orstandard.• Wait for a stable mV to be dis

Strany 10 - TEMPERATURE MEASUREMENTS

18GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GL

Strany 11

19Press the ARROW keys to view the nextcalibration parameter (pressing the key):• The pH calibration slope and offset.The GLP slope is the average o

Strany 12

2Dear Customer,Thank you for choosing a Hanna Instruments product.Please read this instruction manual carefully before using the instruments.This manu

Strany 13

20Notes: • The “OLd” message displayed beside the pH value meansthat this buffer was not used during last calibration. Press2nd then Setup if you want

Strany 14

21SETUPSETUPSETUPSETUPSETUPSetup mode allows viewing and configuring the following parameters:• Calibration Alarm Time Out (pH range only)(dAY)• Custo

Strany 15

22SRETEMARAPPUTESnoitaiverbbAseodtitahWdilaVseulaVtluafeDYAdlavretnitesotretemarapsihtesUaerofebsnoitarbilacneewtebsiegassemeudnoitarbilac.deyalpsidot

Strany 16

23Note: The custom buffers can be set only with 0.01 pH resolution, at25 ºC. Press Range (HI 98150 & HI 98160) or 2nd and Mode(HI 98140) repeatedl

Strany 17

24VIEW LOGGED DATAPress 2nd then RCL to retrieve the informationstored while in measurement mode for thespecific range. “RECALL MEMORY” will bedisplay

Strany 18 - DEAD Probe” tag

25• Press the ARROW keys to change the record number.• Press CAL or Range or CLR to escape from DEL screen and enter viewrecord items mode.Note: Press

Strany 19

26TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATIONTEMPERATURE CALIBRATIONTEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION(for technical personnel only)All

Strany 20

27 mmmmmV CALIBRATION V CALIBRATION V CALIBRATION V CALIBRATION V CALIBRATION (HI 98160 only)(for technical personnel only)All the instruments are

Strany 21 - SETUPSETUP

28PC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEData transmission from the instrument to the PC can be done with theHI 92000 Windows® co

Strany 22 - SRETEMARAPPUTES

29CHR xx Change the instrument range according with the parametervalue (xx):• xx=01 pH range/0.01 resolution• xx=02 pH range/0.1 resolution• xx=03 mV

Strany 23 - LOGGINGLOGGING

3PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONHI 98140, HI 98150 and HI 98160 ar

Strany 24

30• Reading status (2 chars): R - in range, O - overrange, U - under range. First character correspondsto the appropriate range reading. Second charac

Strany 25

31• Rel mV calibration data (if available), which contains:• the calibration offset, with sign (7 chars)• the calibration time, yymmddhhmmss (12 chars

Strany 26 - TEMPERATURE CALIBRATION

32• The mV reading, with sign and decimal point(7 chars) - only for HI 98150 and HI 98160• The logged time, yymmddhhmmss (12 chars)• The calibration o

Strany 27

33If the batteries become weak, the display will flash the battery symbol toadvise the user that approx. 1 hour of working time is left.It is recommen

Strany 28 - PC INTERFACE

34The instrument is provided with the BEPS (Battery Error PreventionSystem) feature, which automatically turns the instrument off when thebatteries le

Strany 29

35• Main active key messages light up for indicating the correspondingactive key.CAL on: CAL key available.CFM blinking: ask confirmation of calibrati

Strany 30

36The temperature limit for the HI 1230B and HI 1618D is from 0 to 80 °Cwith optimum measurements between 20 and 40 °C. The resistance ofglass electro

Strany 31

37PREPARATION PROCEDURERemove the electrode protective cap.DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This isnormal with electrodes and they

Strany 32

38MEASUREMENTVerify if electrode is calibrated before making measurements. Rinse thepH electrode tip with distilled water. Submerse the tip (bottom 3

Strany 33 - BATTERIES REPLACEMENT

39pH CLEANING PROCEDURE• General Soak in Hanna HI 7061 or HI 8061 GeneralCleaning Solution for approximately ½ hour.• Protein Soak in Hanna HI 7073 or

Strany 34 - LCD MESSAGE GUIDE

4When the SMART electrode is attached to the meter again, it isautomatically recognized. This technology allows the operator tooptimize time and effic

Strany 35

40TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDESMOTPMYSMELBORPNOITULOSevissecxe/esnopserwo

Strany 36

41pH CALIBRATION SOLUTIONSHI 50004-01 pH 4.01 Buffer Solution, 20 mL sachet, 10 pcsHI 50004-02 pH 7.01 Buffer Solution, 20 mL sachet, 25 pcsHI 50007-0

Strany 37

42HI 8082 3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL,for double junction electrodesHI 8093 1M KCl+AgCl Electrolyte in FDA approved bottle, 4

Strany 38

43120 mm4.7"12 mm0.5"9.5mm DIA0.37""S" VERSIONHI 1131HI 1131BGlass-body, single junction, refillable, combination pH electrod

Strany 39

44HI 1332BPlastic-body (PES), double junction, refillable, combination pH electrode.Use: general purpose.FC 100BPlastic-body (PVDF), double junction,

Strany 40 - TROUBLESHOOTING GUIDE

45110 mm4.3"12 mm0.5"FC 911FC 911BPlastic-body (PVDF), double junction, refillable with built-in amplifier,combination pH electrode. Use: ve

Strany 41 - ACCESSORIES

46EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&ap

Strany 42 - 4.7"

47RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for theenvironment in which they are used.Operation of th

Strany 43 - 0.25"

48Printed in EUROPE(ROMANIA) MAN98160Y 02/10Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Suppor

Strany 44

5FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTIONHI 98140 & HI 98150HI 98160

Strany 45 - 0.5"

61) Liquid Crystal Display (LCD).2) Range key, to select pH or mV range (HI 98150 & HI 98160only).Setup key, to enter/exit SETUP mode.3) On/Off ke

Strany 46

7EGNARHp0.02ot0.4–Hp00.02ot00.4–Vm9.996±(Vm0002± 05189IH & 06189IH ))Fº0.842ot0.4–(Cº0.021ot0.02–NOITULOSERHp1.0Hp10.0(Vm1.0 05189IH & 06189IH

Strany 47

8OPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEINITIAL PREPARATIONThe instrument is supplied complete with batt

Strany 48 - (ROMANIA) MAN98160Y 02/10

9The LCD will show the pH measurementtogether with the temperature of thesample. In order to take more accuratepH measurements, make sure that theelec