Accessories needed:
Use HI 70031 (1413 µS/cm) EC calibration solution or
HI 70032 (1382 ppm) TDS calibration solution.
Note: The conversion between EC and TDS is made by a built-in
circuit, hence it is requested to calibrate the meter only in EC
or TDS range. The other range is thus automatically calibrated.
PROCEDURE
Pour approximately 4 cm (1½") of a conductivity calibration
solution (e.g. HI 70031) into a beaker. If possible, use plastic
beaker to minimize any EMC interference.
Immerse the probe in the solution.
Wait for a couple of minutes for thermal equilibrium to be
reached.
Tap the probe on the bottom, then shake it lightly while rotating
to make sure no air bubbles remain trapped inside the probe.
Press the µS/cm (or ppm) key.
Turn the EC/TDS calibration knob until
the display shows the EC or TDS reading
at 25°C.
EC/TDSEC/TDS
EC/TDSEC/TDS
EC/TDS
CALIBRATION CALIBRATION
CALIBRATION CALIBRATION
CALIBRATION