Hanna-instruments HI 96721C Uživatelský manuál Strana 17

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 24
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 16
17
CALIBRATION PROCEDURE
Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or
ON/OFF keys.
Warning: Do not calibrate the instrument with standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
Turn the meter on by pressing ON/OFF.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
Press and hold CAL CHECK for three
seconds to enter
calibration mode
. The
display will show “CAL” during
calibration procedure. The blinking
“ZERO” asks for instrument zeroing.
Place the CAL CHECK
TM
Standard
HI 96721-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
Zobrazit stránku 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 23 24

Komentáře k této Příručce

Žádné komentáře