Place the cuvette into the cuvette holder and
ensure that the notch on the cap is positioned
securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the countdown
prior to measurement. An audible “beep”
indicates the end of the countdown period.
Alternatively, wait for 30 seconds then just
press READ/ /TIMER. In both cases, the
lamp, cuvette and detector icons will appear
on the display, depending on the measurement
phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
iron LR on the LCD.
INTERFERENCES
Interference may be caused by:
Cadmium above 4.0 mg/L; Chromium
3+
above 0.25 mg/L; Chromium
6+
above 1.2 mg/L;
Cobalt above 0.05 mg/L; Copper above 0.6 mg/L; Cyanide above 2.8 mg/L; Manganese above
50.0 mg/L; Mercury above 0.4 mg/L; Molybdenum above 4.0 mg/L; Nickel above 1.0 mg/L;
Nitrite ion above 0.8 mg/L.
Sample pH should be between 3 and 4 to avoid developed color to fade or turbidity formation.