Hanna-instruments HI 931100 Uživatelský manuál Strana 11

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 17
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 10
2120
TEMPERATURE CALIBRATION
(for Technical Personnel only)
INITIAL PREPARATION
Prepare a 0°C/32°F bath with ice and water and
another one at 50°C/122°F using hot water at a
known temperature. Place insulation material around
the containers to minimize temperature changes.
Use a
Checktemp
or another accurate thermometer
with a resolution of 0.1° as reference.
Note: A blinking "°C" indicates
that the temperature probe
is not connected.
PROCEDURE
Connect the HI 7662 temperature probe to the
RCA socket on the top of the instrument.
Turn the instrument off.
To enter the temperature calibration mode, press
the CAL and ON/OFF keys simultaneously.
Immerse the temperature probe
into the 0°C/32°F bath together
with the reference thermom-
eter.
When the reading stabilizes,
press the CAL key. The
"0.0°C", "CAL" and "BUF"
symbols will be displayed. The
" " symbol will blink for a
few seconds.
gm
°C
NaCl
0 °C
(32 °F)
°C
50 °C
(122 °F)
°C
°C
0°C
(32°F)
TEMPERATURE
The best working temperature is around 20-25°C
(68-77°F).
For greatest accuracy, it is recommended to cali-
brate the meter with standard solution at a tempera-
ture as close as possible to the temperature of the
sample (±1°C/±2°F).
Na CONCENTRATION WITH HI 931100
It is always possible to determine the Na concentra-
tion by using the formula:
X
Na
= X
NaCl
• (23 / 58.4)
where
X
Na
= concentration of Na
and
X
NaCl
= concentration of NaCl
BUF
CAL
°C
Zobrazit stránku 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 17

Komentáře k této Příručce

Žádné komentáře