THREE-POINT CALIBRATION
(Ratio Turbidity only)
To perform a three-point calibration, continue the procedure with the following steps:
Remove the second standard cuvette.
Place the 100 NTU standard cuvette (or the
third prepared formazin standard) into the holder,
with the cuvette mark aligned to the sign on the
instrument top.
Close the lid and press the Read key. The display
will show the value blinking and the lamp icon
during measurement.
At the end of the measurement, the fourth
proposed calibration point (750 NTU) is
displayed.
At this moment it is possible to exit calibration
by pressing the End key.
If the calibration is terminated, the display will
briefly show “Store...” and the three point
calibration is saved. The instrument returns in the
main screen.
FOUR-POINT CALIBRATION
(Ratio Turbidity only)
To perform a four-point calibration, continue the
procedure with the following steps:
Remove the third standard cuvette.
Place the 750 NTU standard cuvette (or the
fourth prepared formazin standard) into the holder,
with the cuvette mark aligned to the sign on the
instrument top.
Close the lid and press the Read key. The
display will show the value blinking and the lamp
icon during measurement.
At the end of the measurement, the fifth proposed
calibration point (2000 NTU) is displayed.
At this moment it is possible to exit calibration
by pressing the End key.
If the calibration is terminated, the display will
briefly show “Store...” and the four point
calibration is saved. The instrument returns in the
main screen.