IODINE
Iodine
SPECIFICATIONS
Range 0.0 to 12.5 mg/L
Resolution 0.1 mg/L
Accuracy ±0.1 mg/L ±5% of reading at 25 °C
Typical EMC ±0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition
, DPD method. The reaction between iodine and the reagent causes a pink
tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93718-0 DPD Reagent 1 packet
REAGENT SETS
HI 93718-01 Reagents for 100 tests
HI 93718-03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
Select the
Iodine
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cap and add the content of one packet of HI 93718-0
DPD reagent. Replace the cap and shake gently for about 30
seconds to dissolve most of the reagent.
10 mL