Pay attention to push the cuvette completely down in the holder and to align the white point on the
cap to the indexing mark on the meter.
In order to avoid reagent leaking and to obtain more accurate
measurements, close the cuvette first with the supplied HDPE plastic
stopper and then the black cap.
Whenever the cuvette is placed into the measurement cell, it must be
dry outside, and free of fingerprints, oil or dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
Shaking the cuvette can generate bubbles in the sample, causing
higher readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
Do not let the reacted sample stand too long after reagent is added.
For best accuracy, respect the timings described in each specific method.
It is possible to take multiple readings in a row, but it is recommended
to take a new zero reading for each sample and to use the same
cuvette for zeroing and measurement when possible (for most precise
results follow the measurement procedures carefully).
Discard the sample immediately after the reading is taken, or the glass might become permanently
stained.
All the reaction times reported in this manual are at 25 °C (77 °F). In general, the reaction time
should be increased for temperatures lower than 20 °C (68 °F), and decreased for temperatures higher
than 25 °C (77 °F).
INTERFERENCES
In the method measurement section the most common interferences that may be present in an average
sample matrix have been reported. It may be that for a particular treatment process other compounds
do interfere with the method of analysis.