Hanna-instruments HI 83203 Uživatelský manuál Strana 37

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 48
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 36
37
Dissolved Oxygen
Let the sample stand and the flocculent agent will start to settle.
After approximately 2 minutes, when the upper half of the bottle becomes
limpid, add 10 drops of HI 93732C-0.
Replace the cap and invert the bottle until the settled flocculent dissolves completely.
The sample is ready for measurement when it is yellow and completely limpid.
Fill the cuvette up to the mark with 10 mL of the
unreacted (original) sample, and replace the cap. This
is the blank.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0- when the meter is zeroed and ready for
measurement.
Remove the cuvette.
Fill another cuvette up to the mark with 10 mL of the reacted sample and
replace the cap.
Reinsert the cuvette into the instrument.
Press Read to start the reading. The instrument will display the results in mg/L of dissolved oxygen.
INTERFERENCES
Interferences may be caused by reducing and oxidizing materials.
10 mL
Zobrazit stránku 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 47 48

Komentáře k této Příručce

Žádné komentáře