Manganese LR
Insert the second cuvette with the reacted sample into the instrument.
Press Read to start the reading. The instrument displays the results in
µµ
µµ
µg/L of manganese.
Press s or t to access the second level functions.
Press the Chem Frm key to convert the result in µg/L of potassium permanganate (KMnO
4
) and
permanganate (MnO
4
¯).
Press s or t to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO
3
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO
3
Nickel above 40 mg/L
Zinc above 15 mg/L.
Manganese LR
Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent
solution to each cuvette, replace the caps and invert gently
to mix for about 30 seconds.
Add 1 mL of the HI 93748C-0 0.1% PAN indicator
solution to each cuvette, replace the caps and shake
gently.
Add 3 drops of Dispersing Agent (HI 93703-51) to each
cuvette, replace the caps and invert gently to mix for
about 30 seconds.
Place the cuvette with the reacted deionized water (blank)
into the holder and close the lid.
Press Timer and the display will show the countdown prior to zeroing the blank. Alternatively wait for
2 minutes and then press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.