50
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
The temperature user calibration menu can be accessed at the meter startup by pressing simultaniously three
keys as in the below drawing:
Note: The temperature user calibration is performed in three points: around 0°C, 50°C, 100°C.
To start temperature user calibration:
• Insert the probe in the beaker with water at 0°C.
• Press to start the temperature calibration. Adjust the
temperature preset value using
or when
necessary.
• Wait to stabilize and then press to confirm the
calibration point.
• Repeat the above steps for 50°C and 100°C.
• Save the calibration.
• Press to return to measure mode.
Note: Press
if you want to clear the temperature user calibration.